J Neuromonit Neurophysiol > Volume 6(1); 2026 > Article
Chai: Common causes of intraoperative neuromonitoring failure during thyroid surgery and practical solutions: a narrative review

Abstract

Intraoperative neuromonitoring (IONM) provides real-time electromyographic assessment of recurrent laryngeal nerve and external branch of the superior laryngeal nerve function during thyroid surgery, supporting nerve identification, functional verification, and surgical decision-making. Technical and pharmacological failures remain clinically important sources of unreliable monitoring. These failures are distinct from true loss of signal caused by neural injury, and their principal sources include endotracheal tube electrode malposition, residual neuromuscular blockade, equipment malfunction, and procedure-specific difficulties encountered in robotic thyroidectomy and transoral endoscopic thyroidectomy vestibular approach. This narrative review categorizes the recognized causes of technical IONM failure, summarizes available evidence for each, and discusses practical preventive and corrective strategies.

Introduction

Intraoperative neuromonitoring (IONM) has been widely adopted during thyroid and parathyroid surgery, providing real-time functional assessment of the recurrent laryngeal nerve (RLN) and the external branch of the superior laryngeal nerve (EBSLN) [1,2]. The International Neural Monitoring Study Group (INMSG) established a standardized four-step protocol, consisting of V1 (pre-dissection vagus nerve stimulation), R1 (initial RLN stimulation), R2 (post-dissection RLN stimulation), and V2 (post-dissection vagus nerve stimulation), that enables systematic comparison of nerve function before and after thyroid dissection [1]. This protocol helps distinguish global signal loss, which may indicate equipment malfunction or a proximal nerve injury, from focal signal change related to the surgical field.
A key concept in interpreting IONM data is loss of signal (LOS), defined as an electromyographic (EMG) amplitude below 100 µV following an initially satisfactory V1 signal (amplitude˃500 µV), with adequate stimulation at 1-2 mA and after systematic troubleshooting [1,3]. LOS carries prognostic significance for postoperative vocal fold function and directly influences intraoperative strategy, particularly the decision to proceed with contralateral thyroidectomy [3]. However, LOS is a descriptive electrophysiological event, not a diagnosis. It encompasses both true neural LOS, in which actual nerve damage abolishes the EMG response, and false-positive LOS, in which a technical or pharmacological cause produces absent signals despite an intact nerve. When studies report LOS incidence, these two categories are often combined. Gür et al. [4] reported LOS in 5.4% of 456 monitored nerves, but the proportion attributable to technical failure versus true injury varied. Published LOS rates range from approximately 3% to over 20%, reflecting differences in study populations, monitoring protocols, and whether post-troubleshooting or pre-troubleshooting events are counted [3,4].
The present review focuses specifically on the technical and pharmacological causes that produce false-positive LOS or otherwise unreliable EMG signals, collectively termed IONM failure in this article. This is distinct from the separate concept of false-negative monitoring, in which satisfactory intraoperative signals fail to predict postoperative vocal fold dysfunction. False-negative events reflect the inherent predictive limitations of IONM rather than equipment or protocol failure, and their detailed discussion falls outside the scope of this review [1,4].
When IONM fails technically, functional feedback is lost: unnecessary staged operations may follow a false-positive LOS in planned bilateral thyroidectomy, or clinician confidence in the technology may erode after repeated unreliable readings [5]. The causes of such failure are multifactorial, encompassing endotracheal tube (ETT) malposition, residual neuromuscular blockade (NMB), equipment and setting errors, and technique-specific challenges in robotic and transoral endoscopic surgery [3,6]. As a narrative rather than systematic review, this article provides a clinically oriented overview of these common causes and discusses preventive and corrective strategies for each, without attempting a formal meta-analysis or exhaustive evidence grading. The major causes of technical IONM failure and their practical preventive or corrective strategies are summarized in Table 1.

Endotracheal Tube Electrode Malposition

1. Mechanism and clinical relevance

ETT malposition is commonly reported as a leading cause of technical IONM failure [3,6]. Standard IONM relies on surface electrodes embedded on the ETT, which require stable, symmetric contact with both vocal folds. Displacement of the tube moves the electrodes away from the glottic plane, reducing or abolishing the recorded EMG amplitude independently of nerve function, a situation that mimics true LOS.
The most frequently described mechanism is cephalad tube migration during the transition from the neutral intubation position to the neck-extended thyroid surgical position. Tsai et al. [7] measured ETT displacement ranging from 16 mm cephalad to 4 mm caudad after neck extension, with a predominant upward shift. The magnitude depends on individual factors including neck length, cervical spine flexibility, and the degree of extension. Initial tube placement depth, tube size relative to the airway, and fixation technique also influence the likelihood of displacement.
Progressive migration during surgery adds further risk. Loosening of fixation tape, the weight of the anesthesia circuit pulling on the tube, and surgical traction on the trachea can shift the tube incrementally, producing gradual signal degradation that becomes apparent during the later stages of the procedure, particularly during R2 and V2 measurements. Rotational malposition along the tube's longitudinal axis disrupts symmetric electrode-vocal fold contact without necessarily being visible externally.

2. Proposed solutions

Performing tracheal intubation after the patient has been placed in the final neck-extended position eliminates displacement caused by subsequent position changes. Direct laryngoscopy in the extended position typically provides a limited glottic view, but Won et al. [8] demonstrated that video laryngoscopy improved glottic visualization in the thyroid surgical position, supporting its use for this purpose.
After positioning, electrode impedance should be checked (target below 5 kΩ) to confirm adequate electrode-tissue contact [3]. Tube depth should be marked at the incisors, and fixation should be reinforced. If signal quality deteriorates intraoperatively, tube position should be reassessed before any conclusions about nerve function are drawn.

Neuromuscular Blockade Management

1. Impact of neuromuscular blockade on intraoperative neuromonitoring

Nondepolarizing NMB agents such as rocuronium inhibit vocalis muscle depolarization by competitive antagonism at the neuromuscular junction [9]. During complete NMB, recordable EMG responses are absent. Even partial residual blockade reduces EMG amplitudes below diagnostic thresholds, creating unreliable readings that may be misinterpreted as neural LOS. Empis de Vendin et al. [10] reported that 12.5% of patients undergoing monitored thyroidectomy required sugammadex reversal to obtain an adequate V1 signal, though routine reversal was not mandatory in the majority who achieved sufficient spontaneous recovery before nerve stimulation.

2. Rocuronium dose and reversal strategy

The degree and duration of laryngeal muscle blockade correlate with the rocuronium dose administered at induction. Lu et al. [11] compared 0.3 mg/kg with 0.6 mg/kg in a randomized trial and found that the lower dose permitted earlier EMG signal recovery, at the cost of slower onset and a lower rate of immediately satisfactory intubating conditions. The clinical trade-off is between ensuring reliable airway management and minimizing residual NMB at the time of first nerve stimulation; the appropriate choice depends on patient-specific airway factors and the anticipated interval from induction to V1.
When residual NMB prevents adequate signal acquisition, pharmacological reversal may be considered in coordination with the anesthesiology team. The available agents differ in mechanism and profile. Neostigmine, an acetylcholinesterase inhibitor, is inexpensive and widely available. Oh et al. [12] showed in a randomized trial that neostigmine administered immediately after intubation shortened the interval to successful V1 stimulation compared with no reversal, without increasing bucking events. Its limitations include a ceiling effect at deep blockade levels, variable reversal kinetics, and the need for anticholinergic co-administration.
Sugammadex encapsulates rocuronium directly, providing faster and more predictable reversal [13]. Lu et al. [14] confirmed that sugammadex at 2 mg/kg adequately restored EMG signals for IONM in a porcine model. In clinical practice, however, the 2 mg/kg dose, commonly used for moderate NMB reversal in the general anesthetic context, has been associated with intraoperative bucking when applied for early IONM-directed reversal. Chai et al. [15] compared 1 mg/kg with 2 mg/kg in a randomized trial and found that bucking occurred less frequently with the lower dose (13.7% versus 35.3%, p=0.020), while monitoring quality did not differ between groups. These single-center findings suggest that a lower sugammadex dose warrants consideration when the objective is partial reversal sufficient for IONM rather than complete reversal for extubation, though confirmation in broader settings is needed.
The waiting interval before rocuronium re-administration after sugammadex depends on the sugammadex dose, the intended re-dose, and renal function, and should follow current labeling guidance rather than a fixed universal time period [13]. The choice between neostigmine and sugammadex, as well as dose and timing, should be individualized according to neuromuscular monitoring status, the expected interval from induction to V1, institutional practice, anesthetic judgment, and patient-specific factors.

Alternative Laryngeal Recording Electrodes

ETT-based surface electrodes remain the standard recording method for IONM, but their signal quality is inherently sensitive to tube position. Alternative laryngeal electrodes, fixed to the laryngeal framework rather than the ETT, have been investigated as complementary options for reducing position-dependent signal variability in selected cases.
Wu et al. [16] evaluated transcartilage surface recording electrodes (adhesive electrodes placed on the outer surface of the thyroid cartilage) in a porcine model and demonstrated stable EMG recordings with less signal variation during tracheal displacement compared with ETT electrodes. In a clinical series, Jung et al. [17] applied transcartilaginous needle electrodes inserted through the thyroid cartilage and reported higher and more stable EMG amplitudes across all four IONM steps compared with simultaneous ETT recordings, along with improved positive predictive value for LOS detection, though the sample was limited to 54 nerves at risk. In a separate study of 105 nerves, Aygun et al. [18] found that cricothyroid muscle electrodes yielded higher amplitudes for EBSLN monitoring than either thyroid cartilage or ETT electrodes, while thyroid cartilage electrodes performed best for vagus and RLN stimulation.
These results are encouraging, but several limitations temper their clinical applicability. Transcartilaginous needle insertion is an invasive step that carries a risk of local bleeding and becomes technically difficult in patients with calcified thyroid cartilage [17]. Electrode placement is operator-dependent, and no standardized positioning protocol has been widely adopted. The existing evidence derives primarily from single-center studies with moderate sample sizes. Additionally, the added procedural time and cost-effectiveness relative to ETT-based monitoring in routine cases have not been formally evaluated. Alternative electrodes should therefore be viewed as complementary adjuncts for selected situations, such as patients with tracheostomies, small-diameter ETTs, or persistent tube-related signal failure, rather than replacements or routine substitutes for standard ETT-based monitoring.

Equipment and Setting Errors

Equipment-related failures range from disconnected cables to incorrect monitor settings. Individually uncommon, they collectively account for a non-negligible share of clinically relevant troubleshooting episodes and are entirely correctable once identified. Their relative brevity in the literature should not imply low clinical importance; a systematic approach to equipment verification is essential for distinguishing technical artifact from true neural events.
The INMSG troubleshooting algorithm [1,3] recommends a stepwise evaluation when unexpected signal change occurs. The logical starting point is stimulation delivery: the probe tip must be clean, undamaged, and delivering the intended current (1-2 mA for both RLN and vagus nerve stimulation [1]), as contact with electrocautery can compromise probe integrity. Functionality can be verified by stimulating an exposed muscle and observing a visible twitch. If stimulation is confirmed, attention turns to the grounding electrode, since a displaced or poorly conducting ground raises circuit impedance globally and attenuates the recorded signal regardless of nerve status. Recording electrode contact then requires verification: for ETT electrodes, this means reassessing tube position and checking impedance values (target ˂5 kΩ per electrode); for needle or alternative electrodes, insertion depth and wire integrity should be confirmed. Cable connections between the electrodes, junction box, and monitor also warrant physical inspection, since partially disconnected or loosely seated connectors are a frequently overlooked cause of intermittent signal dropout. Finally, monitor settings should be reviewed: the event detection threshold, sensitivity, and stimulation parameters should match the institutional protocol, with the standard LOS criterion for RLN monitoring, an EMG amplitude below 100 µV from an initial satisfactory baseline, serving as a reference [3].
Only after this sequence has been completed without identifying a correctable cause should signal loss be attributed to a neural event. Contralateral vagus stimulation, when available, provides an additional system check: a normal contralateral response in the setting of ipsilateral signal loss supports a true ipsilateral neural event rather than a global equipment failure [1]. Figure 1 summarizes a stepwise approach to troubleshooting unexpected IONM signal loss before attributing the event to true neural injury.

Challenges in Robotic and Transoral Endoscopic Thyroidectomy

Remote-access thyroid surgery, including transaxillary robotic thyroidectomy and transoral endoscopic thyroidectomy vestibular approach (TOETVA), introduces procedure-specific sources of IONM difficulty that extend beyond those encountered through conventional cervicotomy [6,19,20]. The evidence base for monitoring in these approaches is growing but remains more limited than for open surgery, and the considerations described below should be understood as practical precautions rather than established standards.
In robotic approaches, the stimulating probe reaches the target nerve through a longer tissue corridor. Contact between the probe and metal retractors, pooled irrigation fluid, or blood can divert stimulating current away from the nerve, reducing the recorded EMG response. Insulated stimulating probes with a small exposed tip concentrate current delivery. Keeping the operative field dry before each stimulation event further reduces the risk of current dispersion.
TOETVA presents additional challenges related to the proximity of the surgical corridor to the oral cavity and ETT. Saliva tracking along instruments or pooling in the subplatysmal space creates conductive pathways that disperse stimulating current or introduce electrical noise at the recording electrodes [19,20]. Preoperative administration of an antisialagogue such as glycopyrrolate, periodic suctioning of the oral cavity and surgical field, and gauze barriers between the oral and surgical compartments have been described as practical countermeasures [19]. The extent to which these measures eliminate conductive interference has not been rigorously quantified, and awareness of this artifact potential remains important when interpreting equivocal signals during TOETVA.
The longer operative times typical of remote-access procedures also increase the cumulative risk of gradual ETT displacement. Periodic verification of tube position and signal quality, particularly before critical dissection near the RLN, is advisable in these cases. Representative studies supporting the preventive and corrective strategies discussed in this review are summarized in Table 2.

Conclusion

Technical and pharmacological IONM failure during thyroid surgery may produce false-positive LOS or unreliable EMG signals. If unrecognized, these events may lead to unnecessary changes in operative strategy. The principal causes, including ETT electrode displacement, residual NMB, equipment malfunction, and approach-specific signal interference, are individually well described and largely preventable. Intubation after neck extension with laryngoscopic confirmation of electrode position, individualized NMB dosing and reversal, a structured stepwise equipment troubleshooting sequence, and attention to the specific conditions of robotic thyroidectomy and TOETVA together form a practical framework for reducing avoidable monitoring failure. Despite these limitations, ETT-based electrodes remain the current standard recording method for routine thyroid IONM. Alternative laryngeal electrodes may provide supplementary recording options in selected cases, but further standardization and validation are required before broader routine adoption.

Notes

Funding
None.
Conflict of Interest
No potential conflict of interest relevant to this article was reported.
Data Availability
None.
Author Contributions
All work was done by YJC.

References

1. Randolph GW, Dralle H, Abdullah H, Barczynski M, Bellantone R, Brauckhoff M, et al. Electrophysiologic recurrent laryngeal nerve monitoring during thyroid and parathyroid surgery: international standards guideline statement. Laryngoscope 2011;121 Suppl 1:S1-16. doi: 10.1002/lary.21119
crossref pmid
2. Barczyński M, Randolph GW, Cernea CR, Dralle H, Dionigi G, Alesina PF, et al. External branch of the superior laryngeal nerve monitoring during thyroid and parathyroid surgery: International Neural Monitoring Study Group standards guideline statement. Laryngoscope 2013;123 Suppl 4:S1-14. doi: 10.1002/lary.24301
pmid
3. Schneider R, Randolph GW, Dionigi G, Wu CW, Barczynski M, Chiang FY, et al. International Neural Monitoring Study Group guideline 2018 part I: staging bilateral thyroid surgery with monitoring loss of signal. Laryngoscope 2018;128 Suppl 3:S1-17. doi: 10.1002/lary.27359
crossref pmid pdf
4. Gür EO, Haciyanli M, Karaisli S, Haciyanli S, Kamer E, Acar T, et al. Intraoperative nerve monitoring during thyroidectomy: evaluation of signal loss, prognostic value and surgical strategy. Ann R Coll Surg Engl 2019;101:589-95. doi: 10.1308/rcsann.2019.0087
crossref pmid pmc
5. Bae HL, Oh MY, Han M, Wu CW, Chai YJ. Intraoperative neuromonitoring during thyroidectomy does not decrease vocal cord palsy risk, but the cumulative experience of the surgeon may. Surg Today 2024;54:1401-9. doi: 10.1007/s00595-024-02871-5
crossref pmid pmc pdf
6. Dionigi G, Kim HY, Wu CW, Lavazza M, Materazzi G, Lombardi CP, et al. Neuromonitoring in endoscopic and robotic thyroidectomy. Updates Surg 2017;69:171-9. doi: 10.1007/s13304-017-0442-z
crossref pdf
7. Tsai CJ, Tseng KY, Wang FY, Lu IC, Wang HM, Wu CW, et al. Electromyographic endotracheal tube placement during thyroid surgery in neuromonitoring of recurrent laryngeal nerve. Kaohsiung J Med Sci 2011;27:96-101. doi: 10.1016/j.kjms.2010.08.002
crossref pmid pmc
8. Won D, Lee JM, Lee J, Chai YJ, Hwang JY, Kim TK, et al. Usefulness of video laryngoscopy in tracheal intubation at thyroid surgical position for intraoperative neuromonitoring. Sci Rep 2024;14:4980. doi: 10.1038/s41598-024-55537-0
crossref pmid pmc pdf
9. Hemmerling TM, Donati F. Neuromuscular blockade at the larynx, the diaphragm and the corrugator supercilii muscle: a review. Can J Anaesth 2003;50:779-94. doi: 10.1007/BF03019373
crossref pmid pdf
10. Empis de Vendin O, Schmartz D, Brunaud L, Fuchs-Buder T. Recurrent laryngeal nerve monitoring and rocuronium: a selective sugammadex reversal protocol. World J Surg 2017;41:2298-303. doi: 10.1007/s00268-017-4004-9
crossref pmid pdf
11. Lu IC, Tsai CJ, Wu CW, Cheng KI, Wang FY, Tseng KY, et al. A comparative study between 1 and 2 effective doses of rocuronium for intraoperative neuromonitoring during thyroid surgery. Surgery 2011;149:543-8. doi: 10.1016/j.surg.2010.11.006
crossref pmid
12. Oh MY, Chai YJ, Huang TY, Wu CW, Dionigi G, Kim HY, et al. Administration of neostigmine after tracheal intubation shortens time to successful intraoperative neuromonitoring during thyroid surgery: a randomized controlled trial. Sci Rep 2022;12:16797. doi: 10.1038/s41598-022-21282-5
crossref pmid pmc pdf
13. de Boer HD, Driessen JJ, Marcus MA, Kerkkamp H, Heeringa M, Klimek M. Reversal of rocuronium-induced (1.2 mg/kg) profound neuromuscular block by sugammadex: a multicenter, dose-finding and safety study. Anesthesiology 2007;107:239-44. doi: 10.1097/01.anes.0000270722.95764.37
crossref pmid pdf
14. Lu IC, Wu CW, Chang PY, Chen HY, Tseng KY, Randolph GW, et al. Reversal of rocuronium-induced neuromuscular blockade by sugammadex allows for optimization of neural monitoring of the recurrent laryngeal nerve. Laryngoscope 2016;126:1014-9. doi: 10.1002/lary.25577
pmid
15. Chai YJ, Lee JM, Won D, Lee J, Hwang JY, Kim TK, et al. Comparison of sugammadex dose for intraoperative neuromonitoring in thyroid surgery: a randomized controlled trial. Laryngoscope 2021;131:2154-9. doi: 10.1002/lary.29515
crossref pmid pdf
16. Wu CW, Chiang FY, Randolph GW, Dionigi G, Kim HY, Lin YC, et al. Feasibility of intraoperative neuromonitoring during thyroid surgery using transcartilage surface recording electrodes. Thyroid 2018;28:1508-16. doi: 10.1089/thy.2017.0680
crossref pmid
17. Jung SM, Tae K, Song CM, Lee SH, Jeong JH, Ji YB. Efficacy of transcartilaginous electrodes for intraoperative neural monitoring during thyroid surgery. Clin Exp Otorhinolaryngol 2020;13:422-8. doi: 10.21053/ceo.2019.01529
crossref pmid pmc pdf
18. Aygun N, Unlu MT, Kostek M, Caliskan O, Isgor A, Uludag M. Intraoperative cricothyroid muscle electromyography may contribute to the monitorization of the external branch of the superior laryngeal nerve during thyroidectomy. Front Endocrinol (Lausanne) 2023;14:1303159. doi: 10.3389/fendo.2023.1303159
crossref pmid pmc
19. Erol V, Dionigi G, Barczyński M, Zhang D, Makay Ö. Intraoperative neuromonitoring of the RLNs during TOETVA procedures. Gland Surg 2020;9(Suppl 2):S129-35. doi: 10.21037/gs.2019.11.21
crossref pmid pmc
20. Ji YB, Ko SH, Song CM, Sung ES, Lee BJ, Wu CW, et al. Feasibility and efficacy of intraoperative neural monitoring in remote access robotic and endoscopic thyroidectomy. Oral Oncol 2020;103:104617. doi: 10.1016/j.oraloncology.2020.104617
crossref pmid

Figure 1.
Stepwise troubleshooting algorithm for unexpected intraoperative neuromonitoring signal loss. When the EMG signal decreases unexpectedly, technical and pharmacological causes should be assessed systematically before the finding is interpreted as true LOS. Stimulation delivery, circuit and grounding integrity, recording electrode contact, and monitor settings should be checked in sequence. If no correctable cause is identified, contralateral vagus nerve stimulation, when available, can help differentiate global technical failure from an ipsilateral neural event. EMG, electromyography; V1, pre-dissection vagus nerve stimulation; ETT, endotracheal tube; LOS, loss of signal.
jnn-2026-6-1-28f1.jpg
Table 1.
Common causes of technical IONM failure and practical solutions
Cause of IONM failure Mechanism of signal failure Typical intraoperative clue Preventive strategy Corrective strategy
ETT electrode malposition or migration Loss of stable contact between ETT surface electrodes and the vocal folds; cephalad migration after neck extension or progressive displacement during surgery Sudden or gradual EMG amplitude reduction; increased impedance; signal degradation during R2 or V2 despite no obvious nerve trauma Intubate after final neck extension when feasible; use video laryngoscopy for electrode positioning; mark tube depth; reinforce fixation; check impedance after positioning Reassess ETT depth and rotation; confirm electrode-vocal fold contact; suction blood or secretions; reposition the tube before interpreting the event as neural LOS
Residual NMB Suppression of vocalis muscle depolarization despite intact nerve stimulation; reduced EMG amplitude below diagnostic thresholds Weak or absent V1 signal after induction; low EMG amplitude despite adequate stimulation and correct tube position Individualize rocuronium dose; use neuromuscular monitoring; avoid unnecessary redosing before nerve stimulation; allow sufficient recovery before V1 Wait for spontaneous recovery when appropriate; administer neostigmine or sugammadex according to depth of blockade, institutional protocol, and patient factors
Equipment or setting error Failure of stimulation delivery, grounding, recording circuit, cable connection, or monitor configuration Global signal loss; intermittent dropout; inconsistent response across nerves or channels; abnormal impedance Perform pre-use system check; confirm stimulation current, probe integrity, grounding electrode, cables, channel selection, threshold, sensitivity, and filters Check probe function; reconnect or replace cables; reposition ground electrode; verify monitor settings; repeat stimulation after correction
ETT-dependent recording instability Signal variability caused by dependence on tube position, small ETT diameter, difficult airway anatomy, tracheostomy, or persistent poor electrode contact Recurrent amplitude instability despite repeated tube adjustment; inadequate contact or high impedance Select appropriate ETT size when possible; confirm electrode position and impedance before dissection; recognize high-risk airway situations Consider alternative laryngeal recording electrodes, such as transcartilage, transcartilaginous, or cricothyroid muscle electrodes, in selected cases
Robotic thyroidectomy-specific difficulty Longer tissue corridor and indirect stimulation; current dispersion caused by contact with metal retractors, blood, or pooled irrigation fluid Weak or inconsistent response during remote-access stimulation; signal improves after drying field or changing probe contact Use insulated stimulating probes with a small exposed tip; keep the field dry; avoid contact between the probe and metal retractors or pooled fluid Suction blood or irrigation fluid; reapply stimulation under direct visualization; adjust probe angle and contact; repeat stimulation after field optimization
TOETVA-specific difficulty Saliva or oral cavity fluid creates conductive pathways, disperses stimulating current, or introduces electrical noise near recording electrodes Equivocal or unstable signals during transoral procedures; signal variation associated with saliva, pooled fluid, or oral cavity contamination Consider antisialagogue use; perform periodic oral and field suctioning; use gauze barriers between oral and surgical compartments; verify ETT position periodically Remove saliva or pooled fluid; dry the operative field; repeat stimulation after suctioning; reassess ETT position and electrode contact

IONM, intraoperative neuromonitoring; ETT, endotracheal tube; EMG, electromyography; R2, post-dissection recurrent laryngeal nerve stimulation; V2, post-dissection vagus nerve stimulation; LOS, loss of signal; NMB, neuromuscular blockade; V1, pre-dissection vagus nerve stimulation; TOETVA, transoral endoscopic thyroidectomy vestibular approach.

Table 2.
Representative evidence on prevention and correction of IONM failure during thyroid surgery
Topic Study Design Key finding Clinical implication
Standardized LOS definition and troubleshooting Randolph et al. [1]; Schneider et al. [3] International guideline statements LOS should be interpreted only after adequate baseline signal, adequate stimulation, and systematic troubleshooting. Exclude technical and equipment-related causes before diagnosing nerve injury.
Clinical significance of LOS Gür et al. [4] Clinical observational study LOS was reported in monitored nerves and had prognostic value, but technical failure and true neural injury may be mixed in reported LOS events. LOS should be treated as an electrophysiological event requiring interpretation, not as an immediate diagnosis of nerve injury.
ETT displacement after neck extension Tsai et al. [7] Clinical measurement study Neck extension caused ETT displacement, most commonly cephalad migration, which can move surface electrodes away from the vocal folds. ETT position should be confirmed after final thyroid surgical positioning.
Video laryngoscopy in thyroid surgical position Won et al. [8] Clinical study Video laryngoscopy improved glottic visualization in the thyroid surgical position. Intubation or electrode confirmation after neck extension may be facilitated by video laryngoscopy.
Effect of neuromuscular blockade on laryngeal EMG Hemmerling and Donati [9] Review Laryngeal muscles are sensitive to neuromuscular blockade, and residual blockade can suppress EMG responses. Residual neuromuscular blockade should be considered when V1 or early EMG signals are weak or absent.
Selective sugammadex reversal protocol Empis de Vendin et al. [10] Clinical protocol study A subset of monitored thyroidectomy patients required sugammadex reversal to obtain adequate V1 signal. Routine reversal is not always required, but selective reversal may be useful when spontaneous recovery is insufficient.
Rocuronium dose optimization Lu et al. [11] Comparative clinical study Lower-dose rocuronium allowed earlier EMG recovery, but intubating conditions may be less optimal than with higher-dose rocuronium. Rocuronium dosing should balance airway safety with the need for early reliable IONM signals.
Neostigmine-assisted signal recovery Oh et al. [12] Randomized controlled trial Neostigmine after intubation shortened the time to successful IONM signal acquisition. Neostigmine may be considered when residual blockade delays adequate V1 stimulation, depending on institutional protocol.
Sugammadex reversal for IONM Lu et al. [14]; Chai et al. [15] Porcine model and randomized controlled trial Sugammadex can restore EMG signals after rocuronium; lower-dose sugammadex reduced bucking compared with 2 mg/kg while maintaining monitoring quality. Low-dose sugammadex may be considered for IONM-directed partial reversal.
Transcartilage recording electrodes Wu et al. [16] Porcine experimental study Transcartilage surface electrodes produced stable EMG recordings with less variation during tracheal displacement than ETT electrodes. Alternative electrodes may reduce tube-position-dependent signal variability in selected cases.
Transcartilaginous needle electrodes Jung et al. [17] Clinical study Transcartilaginous electrodes produced higher and more stable EMG amplitudes than simultaneous ETT recordings. Transcartilaginous electrodes may be useful when ETT-based signals remain unstable, but invasiveness and standardization remain limitations.
Cricothyroid muscle electrodes for EBSLN monitoring Aygun et al. [18] Clinical study Cricothyroid muscle electrodes produced higher amplitudes for EBSLN monitoring, while thyroid cartilage electrodes performed well for vagus and RLN stimulation. Electrode selection may be tailored to the target nerve and clinical monitoring goal.
TOETVA-specific IONM difficulty Erol et al. [19] Technical review TOETVA introduces unique challenges including current dispersion, saliva-related conductive interference, and difficulty maintaining stable signals. Suctioning, oral cavity control, antisialagogue use, and careful stimulation may reduce signal instability.
Remote-access robotic and endoscopic thyroidectomy Ji et al. [20] Clinical feasibility study IONM was feasible in remote-access robotic and endoscopic thyroidectomy, but standardized monitoring was more challenging than in open surgery. Remote-access approaches require additional attention to stimulation route, electrode stability, and periodic signal verification.

IONM, intraoperative neuromonitoring; LOS, loss of signal; ETT, endotracheal tube; EMG, electromyography; V1, pre-dissection vagus nerve stimulation; EBSLN, external branch of the superior laryngeal nerve; RLN, recurrent laryngeal nerve; TOETVA, transoral endoscopic thyroidectomy vestibular approach.



Editorial Office
Dankook University Hospital, 201 Manghyang-ro, Dongnam-gu, Cheonan 31116, Korea
Tel: +82-41-529-6050   Fax: +82-41-559-7838   E-mail: jnn.kinmos@gmail.com                

Copyright © 2026 by Korean Intraoperative Neural Monitoring Society, Asia-Pacific Society of Thyroid Surgery.

Developed in M2PI